BARC/PUB/2005/0526

 
 

A comparative evaluation of 238Pu determination in NIST-SRM-947-Pu by alpha spectrometry and thermal ionization mass spectrometry

 
     
 
Author(s)

Alamelu, D.; Shah, P. M.; Aggarwal, S. K.
(FCD)

Source

Radiochim. Acta, 2005. Vol. 93: pp. 259-263

ABSTRACT

A comparative evaluation of 238Pu determin-ation in NIST-SRM-947-Pu isotopic reference material by alpha spectrometry and thermal ionization mass spectrometry (TIMS) is presented. Electrodeposited sources with geometric progression (G.P.) method to account for tail   contribution were used in alpha spectrometry. In TIMS, the recently developed approach of using 235U as a monitor isotope in interfering element correction methodology was employed to account for the isobaric interference of 238U at238Pu. The atom % abundance of 238Pu determined by both the methods is in good agreement with each other but is higher than that quoted by NIST in the reference material SRM-947-Pu.

 
 