BARC/PUB/2021/0709

 
 

Determination of boron in in-house graphite reference material by instrumental charged particle activation analysis

 
     
 
Author(s)

Dasgupta, S.; Datta, J.; Swain, K. K.
(ACD)

Source

Journal of Radioanalytical and Nuclear Chemistry, 2021. Vol. 328: pp. 33-38

ABSTRACT

Due to high thermal neutron absorption cross-section, boron (B) content is a binding specification for nuclear grade graphite. Reliable analytical methods are essential to determine boron at trace to ultra-trace levels. Analytical Chemistry Division, Bhabha Atomic Research Centre, India has prepared an in-house graphite reference material (GRM). A Charged particle activation analysis methodology has been developed for quantification of B in graphite matrix. Detection limit for boron using this methodology is 1.4 mg kg−1. This in-house GRM can be used for quantification of B in unknown graphite samples, validating existing analytical methods and for the development of new analytical methods.

 
 