Beta response of CaSO4:Dy based thermoluminescent dosimeter badge and its angular dependence studies for personnel monitoring applications

 
     
 
Author(s)

Munish Kumar; Rakesh, R. B.; Sneha, C.; Ratna, P.; Bakshi, A. K.; Datta, D.
(RP&AD)

Source

Radiation Protection and Environment, 2016. Vol. 39 (3): pp. 132-137

ABSTRACT

Studies on the response of 0.4 mm and 0.8 mm thick Teflon embedded CaSO4:Dy discs are carried out using different beta sources having energy ranging from 0.224 to 3.54 MeV. Angular dependence of the response for 0.4 and 0.8 mm thick thermoluminescence (TL) discs was also studied for 85Kr and 90Sr/90Y beta sources. The ratio of the response of the open disc to that under 1.5 mm thick Perspex filter (DOpen/DPerspex) was estimated for 0.4 and 0.8 mm thick TL discs in the energy range from 0.689 to 3.54 MeV. The evaluation of (DOpen/DPerspex) ratio is necessary as same is used to estimate the energy of beta particles and to apply appropriate correction factor while evaluating beta dose. In addition to above, the optical density and transmission characteristics for 0.4 and 0.8 mm thick Teflon and CaSO4:Dy Teflon embedded discs are also reported.

 
 