Author(s) |
Sahani, G.; Kumar, M.; Sharma, P. K. D.; Sharma, D. N.; Chhokra, K.; Mishra, B.; Agarwal, S. P.; Kher, R. K.
|
Bhabha Atomic Research Centre, Mumbai, India designed and developed a telecobalt unit, Bhabhatron-II. In this paper, the results pertaining to radiation safety of the indigenously developed Bhabhatron-II telecobalt unit are reported. The various tests were carried out as per requirements of International Electrotechnical Commission standards and acceptance criteria. Various devices such as
CaS04:Dy-based thermoluminescent dosimeters, Farmer-type
ionization chamber, water phantom, and radiographic films were used. All the parameters pertaining to radiation leakage/transmission were within the tolerance limits, as per the IEC standard 60601-2-11, except the collimator transmission through X collimators (upper jaw), which marginally exceeds the tolerance limit.
|