BARC/PUB/2016/0034

 
 

Development of a phantom and assessment of 141Ce as a surrogate radionuclide for flood field uniformity testing of gamma cameras

 
     
 
Author(s)

Saxena, S. K.; Kumar, Y.; Malpani, B.; Rakshit, S.; Dash, A.
(IP&AD;RMC)

Source

Applied Radiation and Isotopes, 2016. Vol. 112: pp. 115-121

ABSTRACT

This paper describes an indigenous method for development and deployment of rechargeable liquid filled phantom with newly proposed radionuclide 141Ce for determination of extrinsic uniformity of gamma cameras. Details about design of phantom, neutron irradiation of cerium targets, chemical processing of 141Ce, charging of phantom with 141Ce solution and their performance evaluation are presented. Suitability of 141Ce in quality assurance of gamma cameras used in in-vivo diagnostic imaging procedures has been amply demonstrated.

 
 