Author(s) |
Sumesh, C. G.; Kumar, A. V.; Nair, R. N.; Tripathi, R. M.; Puranik, V. D. (EAD)
|
Source |
Radiation Protection Dosimetry, 2012. Vol. 150 (4): pp. 536-540 |
ABSTRACT
|
Two counting techniques are proposed in this paper to estimate thoron (220Rn) concentration using a Lucas scintillation cell. The alpha activity build-up inside the cell is calculated theoretically by using Bateman equations. The first method is having a minimum detection limit of 325 Bq m-3 and can be used for thoron measurement in thorium-processing plants. In the second method, thoron concentration is calculated using the alpha counts from thoron progenies and is a reference to the first method. The results obtained by these techniques compare well with the double filter method. |
|
|
|