BARC/PUB/2023/0606

 
 

An FPGA-based automated approach for identification and setting of the single electron response threshold voltage in portable TDCR systems

 
     
 
Author(s)

Sharma, M. K.; Agarwal, S.; Kulkarni, M. S.; Kulkarni, D. B.
(HPD;RSSD)

Source

Nuclear Instruments & Methods in Physics Research-A, 2023. Vol. 1048: Article no. 167925

ABSTRACT

The threshold voltage setting of the pulse counting discriminator corresponding to each of the three photomultiplier tubes (PMTs) is crucial in a Triple to Double Coincidence Ratio (TDCR) system for accurate measurement of the activity of a radionuclide. The discriminator threshold voltage of each PMT signal processing channel is required to be set before the single electron response peak and after the thermal noise to allow the detection of a single photoelectron from the photocathode of the PMT. The portable TDCR systems, when subject to different field environments for accurate in-situ activity measurements, may need the re-setting of the discriminator threshold of the PMT channels. However, such miniaturized systems have minimal access to the circuit components for the sake of compactness. Therefore, an automatic solution to obviate the need for access to the circuit components (test connectors/test points) and the measuring instruments in the field is preferred. A Field Programmable Gate Array (FPGA) based automated methodology for identifying and digitally controlling the discriminator threshold voltage in standalone, portable TDCR systems is discussed.

 
 