Deuterium Gas Analysis by Residual Gas Analyzer

 
     
 
Author(s)

Das, B. K.; Shukla, R.; Das, R.; Shyam, A.; Rao, A. D. P.
(E&ED-V)

Source

Journal of Physics-Conference Series, 2012. Vol. 390 (1): pp. 12009

ABSTRACT

Hydrogen gas is generated by electrolysis method in a compact hydrogen generator. A simple procedure reduces handling and storage of hydrogen cylinders for laboratory applications. In such a system, we are producing deuterium gas from heavy water by electrolysis method. After production of the deuterium gas, we have checked the purity level of the outgoing deuterium from the electrolyser. The test was carried out in a high vacuum system in which one residual gas analyser (RGA) was mounted. The deuterium gas was inserted by one manual gas leak valve in to the vacuum system. In this study, the effect of the emission current of the RGA on the detection of the deuterium was performed. In this paper, we will discuss the detail analysis of the deuterium gas and the effect of the emission current on the partial pressure measurement.

 
 