Gupta, Y. K.; Biswas, D. C.; Roy, P.; Nayak, B. K.; Thomas, R. G.; Inkar, A. L.; Vind, R. P.; John, B.; Saxena, A.; Choudhury, R. K.
Source
Nuclear Instruments & Methods In Physics Research-A, 2011. Vol. 629 (1): pp. 149-153
ABSTRACT
Pulse height and timing characteristics of CsI(Tl)-Si(PIN)detector have been investigated for γ-rays and fission fragments(FFs). Energy dependence of the scintillation light output has been studied for FFs from a 252Cf source by degrading their energy in P-10 gas at different pressures. It is seen that the light output for both heavy and light mass groups increases linearly with energy in the interval of 0.2–0.9MeV/A. At a given energy, the light output is more for the heavy mass fragments than for light mass fragments. The time resolution for g-rays was determined to be 134±3 ns using a BaF2 as start and CsI(Tl)-Si(PIN) as stop detector.