BARC/PUB/2018/1346

 
 

Real time implementation of empirical mode decomposition algorithm for ultrasonic nondestructive testing applications

 
     
 
Author(s)

Tarpara, E. G.; Patankar, V. H.
(ED)

Source

Review of Scientific Instruments, 2018. Vol. 89: Article no. 125118

ABSTRACT

Areal time empirical mode decomposition (EMD) algorithm based ultrasonic imaging system has been developed for non- destructive testing (NDT) applications. It is difficult to implement the EMD based signal processing algorithm in real time because it is totally a data-driven process which comprises numerous sifting operations. In this paper, the EMD algorithm has been implemented in the visual software environment. The EMD implementation encompasses two types of interpolation methods: piecewise linear interpolation (PLI) and cubic spline interpolation (CSI). The cubic spline tridiagonal matrix has been solved by using the Thomas algorithm for real time processing. The total time complexity functions for both the implemented PLI and CSI based EMD methods have been computed. For the signal filtering, the partial reconstruction algorithm has been adopted. The baseline correction and noise filtering applications have been presented using an EMD based visual software. The real time practicability and the efficiency of this method have been validated through ultrasonic NDT experimentation for improvement in the time domain resolution of the ultrasonic A-scan raw data. The practical results show that in the noisy environment, it is possible to enhance the signal-to-noise ratio for the visualization and identification of ultrasonic pulse-echo signals in real time. Published by AIP Publishing.

 
 
SIRD Digital E-Sangrahay