Author(s) |
Dash, K.; Thangavel, S.; Rao, S. V.; Chandrasekaran, K.; Chaurasia, S. C.; Arunachalam, J. (CCCM)
|
Source |
Journal of Chromatography-A, 2004. Vol. 1036 (2): pp. 223-227 |
ABSTRACT
|
Trace levels of phosphorus in purified quartz are determined by ion chromatography. In situ reagent purification, matrix digestion and oxidation of phosphorus to orthophosphate ion are carried out simultaneously in a vapour phase digestion (VPD) assembly using a mixture of HF, HNO3 and H2O2. A drastic reduction (475 times) in phosphate blank from reagents (HF/H2O2) was achieved in the VPD through in situ purification of the reagent. The residues remaining after volatilisation (solvent/matrix), mostly consisting of insoluble phosphate/fluorid esalts of divalent and trivalent cations, were solubilised by ion-exchange dissolution. Phosphate was analysed on the IonPac AS17 column with suppressed conductivity detection. The results of the ion chromatography (IC) method were compared with a spectrophotometric method. Accuracy was evaluated by analysing a certified reference material (silicon, NIST 57a). The method detection limit was 0.05µgg−1. |
|
|
|