BARC/PUB/2018/0088

 
 

An automated LS(β)- NaI(Tl)(γ) coincidence system as absolute standard for radioactivity measurements

 
     
 
Author(s)

Joseph, L.; Das, A. P.; Ravindra, A.; Kulkarni, D. B.; Kulkarni, M. S.
(RSSD;DRHR)

Source

Applied Radiation and Isotopes, 2018. Vol. 137: pp. 28-32

ABSTRACT

4πβ-γ coincidence method is a powerful and widely used method to determine the absolute activity concentration of radioactive solutions. A new automated liquid scintillator based coincidence system has been designed, developed, tested and established as absolute standard for radioactivity measurements. The automation is achieved using PLC (programmable logic controller) and SCADA (supervisory control and data acquisition). Radioactive solution of 60Co was standardized to compare the performance of the automated system with proportional counter based absolute standard maintained in the laboratory. The activity concentrations determined using these two systems were in very good agreement; the new automated system can be used for absolute measurement of activity concentration of radioactive solutions.

 
 