BARC/PUB/2017/0437

 
 

Development of X-ray CCD camera based X-ray micro-CT system

 
     
 
Author(s)

Sarkar, P. S.; Ray, N. K.; Pal, M. K.; Baribaddala, R.; Agrawal, A.; Kashyap, Y.; Sinha, A.; Gadkari, S. C.
(TPD)

Source

Review of Scientific Instruments, 2017. Vol. 88 (2): pp. 023702.1-023702.8

ABSTRACT

Availability of microfocus X-ray sources and high resolution X-ray area detectors has made it possible for high resolution microtomography studies to be performed outside the purview of synchrotron. In this paper, we present the work towards the use of an external shutter on a high resolution microtomography system using X-rayCCDcamera as a detector. During micro computed tomography experiments, the X-ray source is continuously ON and owing to the readout mechanism of the CCD detector electronics, the detector registers photons reaching it during the read-out period too. This introduces a shadow like pattern in the image known as smear whose direction is defined by the vertical shift register. To resolve this issue, the developed system has been incorporated with a synchronized shutter just in front of the X-ray source. This is positioned in the X-ray beam path during the image readout period and out of the beam path during the image acquisition period. This technique has resulted in improved data quality and hence the same is reflected in the reconstructed images.

 
 