BARC/PUB/2003/0306

 
 

Development of angle resolved photoelectron spectroscopy beamline at Indus-1 synchrotron radiation source

 
     
 
Author(s)

Das, N. C.; Jha, S. N.; Bhaskara Rao, S. V. N.; Rajasekhar, B. N.; Bhattacharya, S. S.; Sikka, S. K.; Jagannath; Korgaonkar, A. V.; Bhushan, K. G.; Yakhmi, J. V.; Sahani, V. C.
(Spect. Div.;TPPED)

Source

Journal of Optics, 2003. Vol. 32 (1): pp. 27-35

ABSTRACT

A beamline fo angle resolved photoelectron spectroscopy (ARPES) using Indus-I synchrotron source has been recently commissioned at the Centre for Advanced Technology, Indore India. The optical system for the beamline consists of a prefocussing toroidal mirror, 1.4 meter toroidal grating monochromator (TGM-J400) and a post-!ocussing toroidal mirror to cover the spectral region of 40Å to 1000Å. The beamline is coupled with an ARPES experimental station which houses the  indigenously built electron analyser, sample manipulator. LEED/Auger probes, sputter etch Ar ion gun, sample load and lock faciliity. In this paper we are presenting the photoemission spectra of some solid samples which have been reconled  through this beam line.

 
 
SIRD Digital E-Sangrahay