Author(s) |
Das, D.; Ansari, S. A.; Mohapatra, P. K.; Mary, G.; Radhakrishnan, K.; Tripathi, S. C.; Manchnada, V. K. |
Ion-chromatography (IC) as well as high performance
liquid chromatography (HPLC) techniques have
been used as analytical tools for the separation and estimation
of some of the relevant metal ions present in the high
level liquid waste (HLLW). IC was applied for the estimation
of alkali and alkali earth metal ions, viz. Na, Cs, Ba and
Sr using methane sulphonic acid as the eluent on a cation
exchange column. On the other hand, dynamically modified
(with sodium salt of n-octane sulphonic acid) reverse phase
HPLC was followed for the estimation of lanthanides viz.
La, Pr, Nd and Sm using a-hydroxy isobutyric acid as the
eluent on a C-18 column. Sample acidity of 0.01 M HNO3
was optimized for best analytical results. The interferences
of one group of metal ions on the quantification of the other
group of metal ions were studied. The solvent extraction data
(distribution coefficient data) of Na, Cs, Sr, Ba, La, Pr, Nd
and Sm from their mixture was obtained by analyses of the
aqueous samples before and after extraction with extractants
used for actinide partitioning, viz., octyl(phenyl)N,Ndiisobutyl
carbamoyl methylene phosphine oxide (CMPO),
N, N'-dimethyl-N,N'dibutyl tetradecyl malonamide (DMDB
TDMA) and N,N, N', N'-tetraoctyl diglycolamide (TODGA).
The solvent extraction data obtained by IC/HPLC techniques
was compared with those obtained by ICP-AES
technique. A good agreement between the results of the two
techniques validated the present analytical method. |