BARC/PUB/2016/1240

 
 

Optimum Energy Compensation for Current Mode Application of Silicon PIN Diode in Gamma Radiation Detection

 
     
 
Author(s)

Mitra, P.; Srivastava, S.; Singh, S. K.; Akar, D. K.; Patni, H. K.; Topkar, A.; Kumar, A. V.
(RSSD;ED)

Source

IEEE Transactions On Nuclear Science, 2016. Vol. 63: pp. 2777-2781

ABSTRACT

A silicon PIN diode was modeled using the Monte Carlo transport code FLUKA. The code was utilized to identify a composite filter having two or more metals without any perforation, in order to flatten the diode’s absorbed dose rate energy response in the photon energy range 60-1250 keV for DC mode operation. The results were validated experimentally, confirming the utility of the diode for gamma radiation detection and monitoring for wide range of energies.

 
 
SIRD Digital E-Sangrahay