Mathematical Correction for Polyatomic Isobaric Spectral Interferences in Determination of Lanthanides by Inductively Coupled Plasma Mass Spectrometry

 
     
 
Author(s)

Raut, N. M.; Huang, L. -S.; Aggarwal, S. K.; and others
(FCD)

Source

Journal of the Chinese Chemical Society, 2005. Vol. 52: pp. 589-597

ABSTRACT

The determination of lanthanides by Inductively Coupled Plasma Mass Spectrometry (ICP-MS) is com-plicated by several spectral overlaps from M+,MO+ or MOH+ions formed in the ICP. Especially, it is essential to avoid the spectral interferences from lighter lanthanide and Ba polyatomic ions on middle or heavier lanthanides. To tackle this problem, we have developed a mathematical correction method, which reduces all the spectral overlaps from oxide species of Pr, Nd, Ce and Sm over Gd, Tb, Dy and Ho, and Gd, Tb over Yb and Lu. It can also successfully correct the oxide and hydroxide interference of Ba over Eu. The effectiveness of the proposed the mathematical correction scheme is demonstrated for the USGS Standard Rock samples AGV-1 and G-2. The results show that the experimental data obtained by applying the mathematical correction scheme for lanthanides is in good agreement with the reported values, using pneumatic and ultrasonic nebulisation methods, for their ICP-MS analysis

 
 