Author(s) |
Reddy, P. J.; Bhade, S. P. D.; Narayan, K. K.; Narayanan, A.; Babu, D. A. R.; Sharma, D. N.
|
|
Different
methods such as full spectrum DPM (FSDPM), dual isotope estimation
and inclusion methods were studied for the activity quantification
of 3H and 14C in dual beta labeled samples
using liquid scintillation analyzer. A standardized Packard
tritiated water was used as tritium source and 14C
activity was standardized by CIEMAT/NIST method and compared with
the results obtained by the above three methods. Minimum
detectable activity was 2100dpm/l for 3H and 1200dpm/l
for 14C with acounting time of 300min. The accuracy of
the results obtained was found to be within ±10%for 3H: 14C activity ratios
1:1,1:2,2:1,6:1,9:1,13:1and18:1.
|
|
|
|