BARC/PUB/2018/0968

 
 

Development of an ionization chamber for low and medium energy photon dosimetry

 
     
 
Author(s)

Singh, S. K.; Kulkarni, M. S.
(RSSD)

Source

Radiation Protection Dosimetry, 2018. Vol. 182 (2): pp. 215-220

ABSTRACT

Measurement of dose due to low and medium energy X-ray using an ionization chamber needs special considerations as the wall thickness of the detector plays an important role in signal generation. A thin and plane wall ionization chamber having 900 cc volume was developed to study its calibration coefficient with air kerma (Kair) rate and ambient dose equivalent (H*(10)) rate at various X-ray beam qualities generated from a dosimetry grade X-ray machine. Optimized ionization chamber wall thickness was determined through measurements where a flat energy response of the ionization chamber could be established. The measurement shows that the air kerma rate based calibration coefficient for ionization chamber varies up to 45% and ambient dose equivalent rate based calibration coefficient varies up to 20% for wall thickness 1.2–1200 mg/cm2 in the energy range 17.6–213 keV.

 
 
SIRD Digital E-Sangrahay